HI 8314 pH/mV/°C Meter- basic

Manufacturer: HANNA Instruments Inc.

Order number: HI8314

Availability:         5 pcs. in stock
 

Ready to ship today,
Delivery time appr. 1-3 workdays

941.00AED *
 
 
 
 
 

Product information "HI 8314 pH/mV/°C Meter- basic"

  • Pre-amplified pH electrode
  • Easy manual one or two point calibration
  • Automatic Temperature Compensation
  • Low battery indicator

 

HI 8314 is a handheld pH/mV/°C meter designed to be reliable and easy-to-use.

HI 8314 uses the HI 1217D pre-amplified pH electrode and internal temperature sensor.

Manual calibration is performed at one or two points by adjusting the trimmers on the front panel. This instrument is ideal for applications that require a custom calibration point. Manual calibration can be extremely useful in order to achieve better accuracy.

 

Order Information:   HI 8314 is supplied with HI 1217D pH electrode, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 700661 electrode cleaning solution sachets (2), calibration screwdriver, battery, protective case and instructions.

Product Manuals

  Manual: Download

Specification :

Range pH 0.00 to 14.00 pH
  ORP ±1999 mV
  Temperature 0.0 to 100.0°C
Resolution pH 0.01 pH
  ORP 1 mV
  Temperature 0.1°C
Accuracy pH ±0.01 pH
  ORP ±1 mV
  Temperature ±0.4°C (excluding probe error)
pH Calibration   manual, two point, through trimmers (offset ±1 pH; slope: 80 to 110%)
Temperature Compensation   Automatic, 0 to 70°C (32 to 158°F)
Probe   HI 1217D PEI body, pre-amplified pH electrode with internal temperature sensor, DIN connector and 1 m cable (included)
Input Impedance   1012 Ohm
Battery Type / Life   9V / approximately 100 hours of continuous use
Environment   0 to 50°C (32 to 122°F); RH max 100%
Dimensions   145 x 80 x 36 mm (5.7 x 3.1 x 1.4’’)
Weight   230 g (8.1 oz.)

 

 